Vital signs monitor
featuring a color TFT screen, ECG, NIBP, Digital Oximetry, Heart Rate,
Temperature, Respiration, and Nurse Call interface. Optional features
include CO2, CO2/O2
and integrated printer.
Technical Specifications
ECG
| Connectors: |
3- or 5-lead |
| Leads: |
User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I,
II, or III (3-lead) |
| Gain Selection: |
1/2mV, 2mV, 4mV |
| Frequency Response: |
Diagnostic: 0.05 to 100 Hz (-3db); Monitor: 0.5 to 40 Hz (-3db) |
| Sweep Speed: |
12.5, 25, 50mm/sec |
| Electrosurgery Protection: |
Yes
|
| Defibrillator Protection: |
Yes |
| Pacer Detection/Rejection: |
Yes
|
Non-Invasive Blood Pressure
| Technique: |
Oscillometric measure upon inflation |
| Measurement Time: |
<40 sec typical ; 0 to 300mmHg; standard adult cuff |
| Automatic Measurement Cycles: |
2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr |
| Inflation Pressure Range: |
Adult: 0 to 300mmHg; Neonate: 0 to 150 mmHg |
| Resolution: |
1mmHg |
| Accuracy: |
¡Ó2mmHg or ¡Ó2% of reading, which ever is greater, over full
range |
| STAT mode: |
5 min of consecutive readings |
Pulse Oximetry
| Range: |
1% to 99% |
| Resolution: |
1% |
| Accuracy: |
70% to 99% range: ¡Ó2%; 50% to 69% range: ¡Ó3%; <50%: unspecified |
| Method: |
Dual wavelength LED |
Heart (Pulse) Rate
| Source: |
ECG (primary); IBP, Pleth; NIBP |
| Range: |
20 to 300bpm |
| Accuracy: |
¡Ó1bpm or ¡Ó1%, whichever is greater |
Temperature
| Channels: |
2 |
| Range, Accuracy: |
68¢X to 94.8¢X F (20¢X to 34.9¢X C): ¡Ó0.5%; 95¢X to 107.4¢X F (35¢X
to 41.9¢X C): ¡Ó0.2%; 107.6¢X to 113¢X F (42¢X to 45¢X C): ¡Ó0.5% |
| Probe: |
YSI-400 and YSI-700 series |
Respiration
| Rate: |
6 to 150 breaths/min |
| Resolution: |
1 breath/min |
| Accuracy: |
¡Ó1% or ¡Ó1 breath/min, whichever is greater |
| Source: |
User selectable: Smart, CO2, ECG-TTI |
CO2
| Units: |
mmHg; Torr; %; kPa |
| Waveform Scale: |
0 to 3.13%, 0 to 6.25%, 0 to 12.5% and 0 to 25% |
| Method: |
Non-dispersive infrared; autocalibrating |
| N2O Compensation: |
Manual; On/Off, user selectable |
| Range: |
0 to 99mmHg; 0 to 99 Torr; 0 to 12.5%; 0 to 12.5 kPa |
| Resolution: |
1mmHg; 1 Torr; 0.1%; 0.1 kPa |
| Sample Rate: |
Sidestream, selectable: 150ml/min or 50ml/min |
| Accuracy: |
+2mmHg, ¡Ó0.3 volume %, ¡Ó0.3 kPa, or ¡Ó2 Torr at 150ml/min and
respiratory rate of ¡Ó35 breaths/min |
| Rise Time: |
10 to 90%: 350ms at 150ml/min |
O2
| Numeric Display: |
Inspired, expired |
| Method: |
Oxidation-reduction galvanic cell |
| Resolution: |
1% |
| Range: |
0 to 100% |
| Accuracy: |
¡Ó3% volume % |
| Response Time: |
10 to 90%: 600ms at 150ml/min |
Invasive Blood Pressure
| Channels: |
2 |
| Transducer Sites: |
Selectable; ART, PA, LA, RA, CVP, ICP |
| Pressure Range: |
-10mmHg to 300mmHg |
| Zero Offset: |
¡Ó150mmHg |
| Frequency Response: |
0 to 20 Hz |
| Transducer Sensitivity: |
5£gV/V/mmHg |
| Zero: |
Menu-driven |
| Accuracy: |
¡Ó1mmHg or ¡Ó1%, whichever is greater |
| Scale Range: |
-10 to 10, 0 to 20, 0 to 30, 0 to 40, 0 to 60, 0 to 100, 0
to 150, 0 to 200, 0 to 300 |
Alarms
| Indication: |
Audible; visual (EN475) |
| Silence: |
Yes; 2 min or permanent |
Trend Memory
Tabular: ¡@¡@¡@24 hr at 30 sec intervals
Display
| Screen: |
10.4" active color TFT |
| Resolution: |
640 x 480 pixels |
| Waveforms: |
6, maximum |
| Languages: |
English, French, German, Italian, Portuguese, Spanish |
Controls
| Keys: |
9; membrane-activated |
| Rotary Knob: |
Push and rotate; 24 steps/turn |
System Outputs
| Com Ports: |
RS 232-compatible; digital or analog; DB9 (Com 1); mini-din
analog (Com 2) |
| Nurse Call: |
Contact switch; 3.5mm stereo phone jack, standard 14mm length
|
| Defib/Sync: |
BNC connector |
| Video Port: |
High density D-sub 15-pin |
Recorder
| Type: |
Thermal line printer |
| Paper Speed: |
12.5, 25, 50mm/sec |
Mechanical
| Weight: |
Base unit without batteries: 12 lb; 5.45kg |
| Size: |
Base unit: 13.25" (H) x 12.5" (W) x 8.75" (D);
33.65cm (H) x 31.75cm (W) x 22.22cm (D) |
Power Requirements
| Voltage: |
100, 120, 220, 240 VAC; 50/60 Hz |
| Consumption: |
35 W, typical |
| Batteries: |
2; sealed lead acid; life: 2 hr, typical; recharge time: 4.5
hr (Base unit without printer or CO2) |
Environmental
| Operating Temperature: |
59¢X to 95¢X F; 15¢X to 35¢X C |
| Storage Temperature: |
23¢X to 122¢X F; -5¢X to 50¢X C |
| Operating Humidity: |
15% to 90%, noncondensing |
Classification
| Types of Protection: |
Class I equipment |
| Degree of Protection: |
Type CF defib-proof |
| Safety: |
EN 60601-1 |
Quality systems registered to ISO 9001 and CE marking per Annex
II, Clause 3 of Council Directive No. 93/42/EEC concerning medical
devices by TUV Product Service GmbH, Munich.
Specifications subject to change without notice.
|
|