Anesthetic Gas monitor featuring
NDIR technology identifiesand measures five anesthetic agent gases:
Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane. And
Breath-by-breath O2, CO, and N2O
monitoring.Gas flow sampled at 100 ml/min to 200 ml/min. Fast warm-up
time ensures full accuracy within minutes.
Technical Specifications
Gas Monitoring
| Method: |
Sidestream; Non-dispersive infrared (NDIR) |
| Identified Gases: |
Halothane, Isoflurane, Enflurane, Desflurane, Sevoflurane,
CO2, N2O, Oxygen |
Output Data:
|
Inspired and end-tidal gas concentrations; continuous real-time
gas concentrations; respiratory rate (elapsed time since last
breath); agent identification (primary agent and mixed agents);
system information (diagnostic status messages) |
Agent Detection
| Measurement Range: |
| Halothane: |
0 to 10% |
| Isoflurane: |
0 to 10% |
| Enflurane: |
0 to 10% |
| Desflurane: |
0 to 20% |
| Sevoflurane: |
0 to 10% |
| CO2: |
0 to 12.5% |
| N2O: |
0 to 99% |
| Oxygen: |
(Optional) 0 to 100% |
Measurement Accuracy:
| Agents: |
กำ (0.1% abs. + 4% reading) |
| CO2: |
กำ 0.2% abs. or 4% of reading |
| N2O: |
กำ (1.5% abs. + 4% of reading) |
| Oxygen: |
กำ 0.3 vol.% (for 0-90%), กำ 0.4 vol.% (for 91-99%) |
| Time to Detect Agent: |
< 15 seconds @ 200 ml/min |
| Agent Detection Resolution: |
0.1 Volume Percent |
| Mixed Gas Threshold: |
0.2 vol. % +10% of total concentrations |
Rise Time:
| Agents: |
450 msec |
| CO2: |
350 msec |
| N2O: |
400 msec |
| Oxygen: |
600 msec |
Response Time:
| Agents: |
2.50 seconds |
| CO2: |
2.25 seconds |
| NO: |
2.50 seconds |
| Oxygen 1.50 seconds |
| Warm-up Time: |
1 minute to first waveforms; < 20 minutes to full accuracy
|
Respiration Rate
| Range: |
1 to 120 breaths/minute |
| Accuracy: |
กำ 2 bpm or 2% of reading |
System Features
System Configuration:
|
Modular design with by-directional communication,
via cable, to the
Poet Plus 8100 vital signs monitor.
|
Waveform Output:
| Halothane, Isoflurane, Enflurane, Desflurane, Sevoflurane,
CO2, N2O, O2
|
| Concentration Units: |
Vol.%, Torr, kPa, mmHg |
| Flow Rates: |
100 ml/min, 150 ml/min, 200 ml/min |
| Occlusion Clearing: |
Automatic |
| Auto Zeroing: |
Occurs 30 to 60 minutes |
| Duration: |
3.0 to 7.0 seconds |
| Manual user calibration not required. |
| Temperature stabilized optical assembly |
Alarm Characteristics:
| EN 475, Adjustable; with audible and visual indications
from the vital signs monitor. |
| Alarm Levels: |
High, Medium, Low, Informational |
Alarm Modes: Adult/Pediatric/Neonate
High and low limit settings for each mode.
Trends:
| 24 hours of stored data in vital signs monitor
with graphical and tabular trends. |
Power Requirements
| 10 to 28 VDC: |
9 watts warm-up, 6 watts operational |
| Receives power from vital signs monitor. |
Mechanical
| Weight: |
6.5 lbs (2.95 kg) |
| Size: |
3.8¨ (H) x 10.4¨ (W) x 8.2¨ (D); 9.65 cm (H) x
26.4 cm (W) x 20.8 cm (D) |
Environmental
| Operating Temperature: |
15° to 35° C; 59o° to 95° F |
| Storage Temperature: |
-5° to 50° C; 23° to 122° F |
| Operating/Storage Humidity: |
15% to 95%, noncondensing |
| Altitude: |
-1,000 to 10,000 ft. (-300m to 3000m) |
| Type of Protection: |
Class I Equipment |
| Degree of Protection: |
Type CF, Defibrillator-proof |
| Protection against ingress: |
Ordinary |
Quality systems registered to ISO 9001 and CE marking per Annex
II, Clause 3 of Council Directive No. 93/42/EEC concerning medical
devices by TUV Product Service GmbH, Munich.
Specifications subject to change without notice.
|
|