Download |
| Poet IQ2 8500Q Anesthetic Gas Monitor |
8500Q Poet IQ2 |
| 8500Q Poet IQ2 Anesthetic Gas Monitor - featuring automatically
NDIR technology identifies and quantifies inspired and expired CO2,
N2O, O2 and five anesthetic
agents. Five anesthetic agent gases: Halothane, Isoflurane, Sevoflurane,
and Desflurane. |
|
|
Regional Manager |
|
Please contact: |
| ¡@ |
Tel:
+886-3-5781410 |
| Features and Benefits |
Mr. Jone Min (Twn.) |
| ¡@ |
EXT. 6830 |
State-of-the-art NDIR technology identifies and measures
five anesthetic agent gases:
Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane. |
|
|
Customer Service |
| ¡@ |
Please contact: |
| Automatic and manual primary agent detection. |
Tel:
+886-3-5781410 |
| ¡@ |
Mr. Anderson Chen |
| Breath-by-breath CO2, O2,
N2OO monitoring. |
EXT. 6802 |
| ¡@ |
¡@ |
| Gas flow sampled at 100 ml/min to 200 ml/min. |
¡@ |
| ¡@ |
¡@ |
| Mixed agent identification. |
¡@ |
| ¡@ |
¡@ |
| Measurement accuracy is not affected by alcohol or metabolic
ketones. |
¡@ |
| ¡@ |
¡@ |
| Fast warm-up time ensures full accuracy within minutes. |
¡@ |
| ¡@ |
¡@ |
| No user calibration required; minimal user interaction. |
¡@ |
| ¡@ |
¡@ |
| Intuitive, easily configured user interface minimizes
training and is ideal for multiple users. |
¡@ |
| ¡@ |
¡@ |
| Bright, real-time display of numerical values and waveforms
provides instant notification of changing patient status. |
¡@ |
| ¡@ |
¡@ |
| Lightweight, portable design provides flexible workspace
options. |
¡@ |
| ¡@ |
¡@ |
| Technical Specifications |
¡@ |
| ¡@ |
¡@ |
| GAS MONITORING |
¡@ |
| Method: |
Sidestream; Non-dispersive infra-red
(NDIR) |
| Identified Gases: |
Halothane, Isoflurane, Enflurane, Desflurane,
Sevoflurane, CO2, N2OO,
Oxygen |
| Concentration Units: |
Vol%, Torr, kPa, mmHg |
| Flow Rates: |
100 ml/min, 150 ml/min, 200 ml/min |
| Occlusion Clearing: |
Automatic |
| Auto Zeroing: |
Occurs 30 to 60 minutes |
Duration:
|
3.0 to 7.0 seconds
Manual user calibration not required.
Temperature stabilized optical assembly. |
|
¡@ |
| ¡@ |
¡@ |
| AGENT DETECTION |
¡@ |
| Measurement Range: |
¡@ |
| Halothane: |
0 to 10% Isoflurane: 0 to 10% |
| Enflurane: |
0 to 10% |
| Desflurane: |
0 to 20% |
| Sevoflurane: |
0 to 10% |
| CO2: |
0 to 12.5% |
| N2OO: |
0 to 99% |
| Oxygen: |
0 to 100% |
| Measurement Accuracy: |
Agents:¡Ó(0.1% abs. + 4% reading) |
| CO2: |
¡Ó0.2% abs. or 4% of reading |
| N2O: |
¡Ó(1.5% abs. + 4% of reading) |
| Oxygen: |
¡Ó0.3 vol% (for 0-90%), ¡Ó0.4 vol% (for
91-99%) |
| Time to Detect Agent: |
< 15 seconds @ 200ml/min |
| Agent Detection Resolution: |
0.1 Volume Percent |
| Mixed Gas Threshold: |
0.2 vol. % +10% of total concentrations |
| Rise Time: |
Agents: 450 msec |
| CO2: 350 msec;
N2O: |
400 msec; Oxygen: 600 msec |
| Response Time: |
Agents: 2.50 seconds |
| CO2: |
2.25 seconds |
| N2O: |
2.50 seconds |
| Oxygen 1.50 seconds |
¡@ |
|
¡@ |
| ¡@ |
¡@ |
| RESPIRATION RATE |
¡@ |
| Range: |
1 to 120bpm |
| Accuracy: |
¡Ó2bpm |
|
¡@ |
| ¡@ |
¡@ |
| SYSTEM FEATURES |
¡@ |
| ¡@ |
¡@ |
| Display |
¡@ |
| Main Display: |
Active color TFT |
| Size: |
5.9" Display Area |
Resolution: |
Internal Display; 320 x 240 pixels
External Video; 640 x 480 pixels
Number of Traces: 3 maximum |
|
¡@ |
| ¡@ |
¡@ |
| Controls |
¡@ |
| Keys: |
7; membrane-activated |
| Rotary knob: |
Push and rotate; 24 steps/turn |
|
¡@ |
| ¡@ |
¡@ |
| SYSTEM OUTPUTS |
¡@ |
| Com 2; Serial/Analog, mini-DIN8 |
| Com Ports: |
Com 1; RS 232-compatible, Serial DB-9 |
| Video Port: |
Color VGA compatible, High Density
D-Sub 15 |
|
¡@ |
| ¡@ |
¡@ |
| TRENDS |
¡@ |
| Memory: |
24 hours of tabular & graphic patient
data |
|
¡@ |
| ¡@ |
¡@ |
| POWER REQUIREMENTS |
¡@ |
| Voltage: |
100, 120, 220, 240 VAC; 50/60 Hz |
| Consumption: |
40 W, typical |
|
¡@ |
| ¡@ |
¡@ |
| MECHANICAL |
¡@ |
| Weight: |
13 lbs. (5.9 kg.) |
Size: |
6.5" (H) x 11.0"(W) x 12.0"(D);
16.5cm (H) x 27.94cm (W) x 30.48cm (D) |
|
¡@ |
| ¡@ |
¡@ |
| ENVIRONMENTAL |
¡@ |
| Operating Temperature: |
15¢J to 35¢J ; 59¢K to 95¢K |
| Storage Temperature: |
-5¢J to 50¢J ; 23¢K to 122¢K |
| Operating/Storage Humidity: |
15% to 95%, noncondensing |
| Altitude: |
-1,000 to 10,000 ft. (-300m to 3000m) |
| Type of Protection: |
Class I Equipment |
| Degree of Protection: |
Type CF |
| Protection against ingress: |
Ordinary |
|
¡@ |
| ¡@ |
¡@ |
Compatible with external VGA displays.
External display not included. |
¡@ |
| ¡@ |
¡@ |
| Quality systems registered to ISO 9001 and CE marking
per Annex II, Clause 3 of Council Directive No. 93/42/EEC concerning
medical devices by TUV Product Service GmbH, Munich. Specifications
subject to change without notice. |
¡@ |
| |
¡@ |
| ¡@ |
¡@ |