504DX/P Portable Pulse Oximeter 504DX Portable Oximeter
504DX Portable Pulse Oximeter - Portable monitor featuring digital oximetry for continuous or spot-check monitoring. 24-hour internal memory supports Profox Oximetry sleep study software.

Technical Specifications

Pulse Oximetry
Range: 0% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ¡Ó2%
40% to 69% range: ¡Ó3%
<40%: unspecified
Method: Dual wavelength LED

Heart (Pulse) Rate
Range: 20 to 300bpm
Resolution: 1bpm
Accuracy: ¡Ó1bpm or ¡Ó1%, whichever is greater

Alarms
Physiological/System: Audible; visual (EN475); adjustable volume
Silence: Yes; Temporary (2 min) or Permanent
Parameter/Range/ Default: High Saturation: 70% to 98% or Off; Off
  Low Saturation: 30% to 99% or Off; 85%
¡@ High Pulse: 80 to 250bpm or Off; 150bpm
  Low Pulse: 20 to 150bpm or Off; 50bpm

Displays
Type: LED

Parameters:

SpO2; pulse rate; pulse bar
Status: Alarm Silence; Sensor; Low Battery; External Power; Neonatal Mode
Text: Alphanumeric LCD
Languages: Selectable; English, French, German, Italian, Portuguese, Spanish

Data Storage
Capacity: ¡@24 hr at 4 sec intervals

System Output
Com Port: ¡@¡@¡@¡@RS 232-compatible; digital or analog; mini-din
 
Serial Interfaces: ¡@ASCII, text, or CSV; Profox-compatible


Printer

Type: Thermal line printer; optional
Tabular Mode: 5, 15, 30 sec intervals; 1, 2, 3, 5, 10, 15, 30, 60 min intervals; Off
Graphical Mode:
Resolution: 5, 15, 30 sec; 1, 2 min
Span: 1, 4, 8, 12, 24 hr

Mechanical
Weight: With printer and battery: 3.34 lb; 1.52kg
Size: Base unit: 5.67" (H) x 7" (W) x 4.81" (D); 14.4cm (H) x 17.8cm (W) x 12.2cm (D)


Power Requirements

Voltage: 60 Hz: 110/120 VAC; 50 Hz: 220/240 VAC
Battery: Sealed lead acid; life: 6 hr, typical; recharge time: 8 hr

Environmental
Operating Temperature: 32° to 113° F; 0° to 45° C
Storage Temperature: 23° to 131° F; -5° to 55° C
Operating Humidity: 15% to 90%, noncondensing

Approved by UL, CSA, CE. Quality systems registered to ISO 9001 and CE marking per Annex II, Clause 3 of Council Directive No. 93/42/EEC concerning medical devices by TUV Product Service GmbH, Munich.

Specifications subject to change without notice.