506DXN2 Series Vital Signs Monitor
- Vital signs monitor featuring Digital Oximetry and NIBP.
Optional IVACR predictive temperature also available.
Custom Configurations:
| 506DXN2: |
SpO2 and NIBP |
|
| 506DXNP2: |
SpO2, NIBP and printer |
|
| 506DXNT2: |
SpO2, NIBP, and predictive temperature |
|
| 506DXNTP2: |
SpO2, NIBP, predictive temperature,
and printer |
Technical Specifications
Non-Invasive Blood Pressure
| Technique: |
Oscillometric measure upon inflation |
|
| Measurement Time: |
<40 sec typical; 0 to 300mmHg; standard adult cuff |
|
| Automatic Measurement Cycles: |
1, 2, 3, 5, 10, 15, 30, 45, 60 min; 2, 4 hr; STAT |
|
| Inflation Pressure Range: |
0 to 300mmHg (Adult); 0 to 150mmHg (Neonate) |
|
| Resolution: |
1mmHg |
|
| Accuracy: |
¡Ó2mmHg or ¡Ó2% of reading, whichever is greater, over full
range |
|
| STAT mode: |
5 min of consecutive readings |
Pulse Oximetry
| Range: |
1% to 99% |
|
| Resolution: |
1% |
|
| Accuracy: |
70% to 99% range: ¡Ó2%; 40% to 69% range: ¡Ó3%; <40%: unspecified |
|
| Method: |
Dual wavelength LED |
|
| Audio: |
Tone with each detected pulse; pitch varies with saturation |
Heart (Pulse) Rate
| Range: |
20 to 300bpm |
|
| Accuracy: |
Greater of ¡Ó1bpm or ¡Ó1%, of reading unspecified for <30
or > 240 bpm |
|
| Update Time: |
Every sec |
|
| Resolution: |
1bpm |
|
| Pulse Volume: |
10 levels or Off |
Temperature
| Type: |
IVACR |
|
| Method: |
Predictive |
|
| Channels: |
1 |
|
| Range: |
Monitor mode: 80.0° to 108.0° F; 26.7° to 42.2°
C |
|
| Accuracy: |
¡Ó0.2° F (¡Ó0.1° C); tested in calibrated water bath |
|
| Resolution: |
0.1° F or 0.1° C |
|
| Probe/Cable: |
Oral: Cat. No. 444; rectal (optional): Cat. No. 445 |
Alarms
| Indication: |
Audible; visual |
|
| Volume: |
10 levels or Off |
|
| Silence: |
Yes; 2 min or permanent |
|
| Parameter/Range/Default: |
| High Saturation: |
70% to 98% or Off; Off |
|
| Low Saturation: |
1% to 99% or Off; 90% |
|
| High Pulse Rate: |
80 to 250bpm or Off; 150bpm |
|
| Low Pulse Rate: |
20 to 150bpm or Off; 50bpm |
|
| High Systolic: |
75 to 240mmHg or Off; 200mmHg |
|
| Low Systolic: |
50 to 150mmHg or Off; 50mmHg |
|
| High Diastolic: |
50 to 180mmHg or Off; 100mmHg |
|
| Low Diastolic: |
15 to 50mmHg or Off; 30mmHg |
|
| High Mean: |
70 to 200mmHg or Off; 150mmHg |
|
| Low Mean: |
25 to 125mmHg or Off; 50mmHg |
|
| High Temperature: |
80.6° to 107.6° F (27° to 42° C) or Off; 100°
F (37.8°oC) |
|
| Low Temperature: |
80.6° to 107.6° F (27° to 42° C) or Off; 94°
F (34.4° C) |
Trend Memory
Memory: ¡@¡@ ¡@Last 24 hrs and 250 readings: SpO2,
pulse rate, NIBP, temperature
Displays
| Type: |
LED |
|
| Parameters: |
NIBP, SpO2; temperature, pulse rate;
pulse bar |
|
| Status: |
Alarm Silence; Sensor; Low Battery; External Power; Neonatal
Mode |
|
| Text: |
Alphanumeric LCD |
|
| Languages: |
Selectable; English, French, German, Italian, Portuguese,
Spanish. |
System Output
| Com Port: |
RS 232-compatible; digital or analog; mini-din |
Printer
| Type: |
Thermal array |
|
| Tabular Mode: |
1, 2, 5, 10, 15, 30, 60 min intervals; 2, 4, 8, 12, 24 hr
intervals; Auto BP; Off |
Mechanical
| Weight: |
Base unit with battery: 4.4 lb; 2kg |
|
| Size: |
Base unit: 6.5" (H) x 7" (W) x 6.125" (D);
16.5cm (H) x 17.8cm (W) x 15.5cm (D)
|
¡@¡@¡@
Power Requirements
| Voltage: |
60 Hz: 110/120 VAC; 50 Hz: 220/240 VAC |
|
| Battery: |
Sealed lead acid; life: 6.5 hr with NIBP every 15 min; recharge
time: 6 hr
|
Environmental
| Operating Temperature: |
32° to 113° F; 0° to 45° C |
|
| Storage Temperature: |
23° to 122° F; -5° to 50° C |
|
| Operating Humidity: |
15% to 90%, noncondensing |
Classification
| Type of Protection: |
Class II equipment |
|
| Degree of Protection: |
Type CF defibrillator-proof equipment |
Approved by UL, CSA, CE.
Quality systems registered to ISO 9001 and CE marking per Annex
II, Clause 3 of Council Directive No. 93/42/EEC concerning medical
devices by TUV Product Service GmbH, Munich.
Specifications subject to change without notice.
.
|
|