large screen
Vital signs monitor with optional high-resolution TFT color display
that combines all essential parameters in one compact system: ECG,
pulse oximetry, NIBP, temperature and respiration.
Technical Specifications
ECG
| Number of Leads: |
3 |
|
| Lead Selection: |
I, II, III |
|
| Gain Selection: |
0.5 mV, 1 mV, 2 mV, 4 mV, 8 mV full scale |
|
| Frequency Response: |
Diagnostic: 0.05 to 100 Hz Monitor: 0.5 to 40 Hz |
|
| Isolation: |
Conforms to IEC 601-1 |
|
| Sweep Speed: |
25, 50 mm/sec |
|
| Electrosurgery Protection: |
Yes |
|
| Defibrillation Protection: |
Yes |
|
| Pacer Rejection: |
Yes |
Heart Rate Display and Alarms
| Range: |
30 to 300 BPM (ECG, Sp02)
|
|
| Accuracy: |
> of ¡Ó1 BPM or 1% |
|
| Alarm Indication: |
Audible, Visual |
|
| Alarm Volume: |
Adjustable: 10 settings + Off, user-selectable |
|
| Alarm Silence: |
Yes, 2 minutes |
Noninvasive Blood Pressure
| Technique: |
Automatic oscillometric upon inflation |
|
| Max. Inflation Pressure: |
Adult 300mmHg; Neonate 150mmHg |
|
| Measurement Cycle: |
< 40 sec. average (0 to 300mmHg: standard adult cuff) |
|
| Automatic Measurement Cycles: |
1, 2, 3, 5, 10, 15, 30, and 60 min; STAT pressure interval
is 5 min
|
Impedance Respiration
| Range: |
9 to 150 breaths/minute |
|
| Resolution: |
1 breath/minute |
|
| Accuracy: |
> of ¡Ó1% or ¡Ó1 breath/minute |
Temperature
| Range: |
20 to 45¢X C, 68 to 113¢X F |
|
| Accuracy: |
¡Ó 0.1¢X C |
|
| Display Resolution: |
0.1¢X C or 0.1¢X F |
|
| Probe: |
YSI-400 Series |
Oxygen Saturation (Sp02)
| Averaging Time: |
Selectable 3 to 21 seconds |
|
| Search Time: |
Selectable, 10 to 40 seconds |
|
| Waveform: |
User Selectable |
|
| Gain: |
User Selectable |
|
| Sensor Type: |
Reusables: DOX™, MultiSite™, SHELL™; Disposables:
Adult, Pediatric, Infant, Neonatal |
|
| Audio: |
Beep with each detected pulse, pitch varies with oxygen saturation
|
|
| Range: |
1 to 99% |
|
| Resolution: |
1% |
|
| Accuracy: |
¡Ó2% (70 to 99%); ¡Ó3% (50 to 70%) |
|
| Method: |
Dual wavelength LED |
External Power Supply
| nput Voltage: |
I90-260 VAC; 47-63 Hz |
|
| Output Voltage: |
16V DC |
Operating Environment
| Temperature: |
15¢X to 40¢X C (59¢X to 104¢X F) |
|
| Humidity: |
15 to 90% (non-condensing) |
|
| Storage Temperature: |
-5¢X to 50¢X C (23 to 122¢X F) |
Display
| Main Display: |
Active color LCD or monochrome |
|
| Size: |
4.76" (121mm) W x 3.58"¨ (91mm) H |
|
| Resolution: |
320 pixels W x 240 pixels H |
|
| Numerical Display: |
LED (green: Sp02%; red: Rate, Sys,
Dia, Map) |
|
| Number of Traces: |
3 maximum |
System Outputs
| Digital: |
DB-9 (serial interface) |
|
|
| Analog Output: |
BNC, ¡Ó1V range |
Trends
Memory: ¡@¡@¡@ ¡@ ¡@ ¡@ 24 hours of tabular patient data at 3-minute NIBP
interval and 1 minute other parameters
Battery
| Type: |
Rechargeable lead acid |
|
| Life: |
1 hour |
|
| Recharge Time: |
6 hours to 100% capacity |
Physical Properties
| Size: |
19 cm (7.6¨¨) H x 26 cm (10.125¨)
W x 25 cm (9.75¨) D |
|
| Weight: |
3.83 kg (8.5 lbs) |
Classifications
| Type of Protection: |
Class I equipment |
|
| Degree of Protection: |
Type CF defibrillator-proof equipment |
|
| Protection Against Liquids: |
IPX1, drip-proof equipment
|
|
|